- 95 -
•
Put reference weight (as 100%) on the pan, wait for the indication to be
stable (
symbol is displayed) and confirm the weight by pressing
button.
•
The software automatically enters the measured load value as a reference
weight and enters <
PERCENT WEIGHING
> mode, 100.000% value is
displayed within the main section, reference sample weight value (target
weight) is displayed in the bottom line (if such option has been selected for
the function <
INFORMATION
>)
17.5.2 Entering reference weight
Procedure:
•
Press F button to which function <
ENTER SAMPLE
> is assigned,
<
ENTER SAMPLE
> editing window is displayed.
•
Using navigating buttons to enter the known reference weight.