- 142 -
UG – give current weight unit
Command overview:
Command gives current unit.
Format:
UG <CR><LF>
Accessible responses:
UG_x_OK<CR><LF>
– command carried out, current unit given as a
response
UG_I <CR><LF>
– command understood but not accessible at this
moment
x
– parametr, unit
Example:
Command:
UG<CR><LF>
– give current unit
Response:
UG_ct_OK<CR><LF>
– currently selected unit is „ct”
BP - Activate sound signal
Command overview:
Command activates BEEP sound signal for a specified amount of time.
Format:
BP_TIME <CR><LF>
Accessible responses:
BP_OK <CR><LF>
- command carried out, BEEP sound signal activated
BP_E” <CR><LF>
- no parameter or incorrect format
BP_I <CR><LF>
- command understood but not accessible at this
moment
TIME
– parameter, decimal specifying how long shall the sound last, parameter
given in [ms]. Recommended range <50 - 5000>.
If value greater than the permissible high limit is given, than BEEP sound is
operated for the maximum permissible amount of time.
Example:
Command:
BP_350<CR><LF>
- activate BEEP for 350 ms
Response:
BP_OK<CR><LF>
- BEEP activated
CAUTION! – BEEP sound activated via BP command is inhibited if in-course of its
activation the sound is activated by means of other device: keypad, touch panel,
proximity sensors.
Send all the implemented commands
Format:
PC CR LF