- 146 -
ARS_OK <CR><LF>
- command carried out
ARS_E <CR><LF>
- an error occurred on command carrying out, no
parameter available or incorrect format
ARS_I <CR><LF>
- command understood but not accessible at this
moment
n
– parameter, decimal value determining value release options
n →
1 – fast
2 – fast + reliable
3 – reliable
Caution:
The numbering is assigned to a particular value release option and it is identical for
all scale types.
The command changes settings for a current working mode if, for a particular scale
type, value release settings are assigned to the working mode.
Example:
Command:
ARS_2 <CR><LF>
– set value release parameter to fast+reliable option
Response:
ARS_OK <CR><LF>
– fast+reliable option has been set
Set last digit
LDS – (Last Digit, Set)
Command overview:
Command lets the user set last digit for a particular device.
Format:
LDS_n <CR><LF>
Accessible responses:
LDS_OK <CR><LF>
- command carried out
LDS_E <CR><LF>
- an error occurred on command carrying out, no
parameter available or incorrect format
LDS_I <CR><LF>
- command understood but not accessible at this
moment
n
– parameter, decimal value determining last digit settings
n →
1 – always
2 – never
3 – when stable
Caution:
The numbering is assigned to a particular last digit option and it is identical for all
scale types.
The command changes settings for a current working mode if, for a particular scale
type, last digit settings are assigned to the working mode.