- 139 -
OMI - Give available working modes
Command overview:
Command returns accessible working modes.
Format:
OMI <CR><LF>
Accessible responses:
OMI <CR><LF>
n_”Mode name” <CR><LF>:
n_”Modename” <CR><LF>
OK <CR><LF>
- command carried out, accessible working
modes returned
OMI_I <CR><LF>
- command understood but not accessible at
this moment
Mode name
– parameter, working mode name, inserted in between inverted
comas. The name takes form given on a display, it is provided in a currently
selected language.
n
– parameter, decimal value determining working mode number
n →
1 – Weighing
2 – Parts Counting
3 – Deviations
4 – Dosing
5 – Formulas
6 – Animal Weighing
8 – Density of Solid Bodies
9 – Density of Liquids
10 – Peak Hold
11 – Totalizing
12 – Checkweighing
13 – Statistics
14 – Pipettes Calibration
CAUTION: Working modes numbering is identical for each kind of scale. The
numbers are assigned to working modes names. Some scales give only the
number as a response.
Example 1
:
Command
:
OMI <CR><LF>
- give accessible working modes
Response
:
OMI <CR><LF>
- accessible working modes are given;
2_”Parts Counting” <CR><LF>
name
4_”Dosing” <CR><LF>
12_”Checkweighing” <CR><LF>
OK <CR><LF>
- command carried out
Example 2
:
Command
:
OMI <CR><LF>
- give accessible working modes