Test results:
*Deviation for Max.
*Value of repeatability of indication expressed as standard deviation
An exemplary report:
..........................................................
--------- GLP Autotest: Report ---------
Balance type
PS 3000.R2
Balance ID
400010
User
Admin
Software rev.
v.0.4.9
Date
2013.07.16
Time
09:17:16
-----------------------------------------------
Number of measurements
10
Reading unit
0.001/0.01 g
Internal weight mass
1402.094 g
Filter
Average
Value release Fast &Reliable
-----------------------------------------------
Deviation for Max.
-0.118 g
Repeatability
0.0088 g
Signature
…………..........................................
PROCEDURE:
Press
button to run parameter <P6.9 GLP AUTOTEST>.
The software starts GLP autotest procedure, the procedure is carried out
automatically, no operator interference is required till the end.
A user can stop the procedure at any time by pressing
button of the GLP
process window.
Once the procedure is completed, the value of standard deviation of all
measurements is shown in the main display, and message <RESULT> in the
bottom line, the final report is printed out automatically (see the exemplary report
above). It is possible to print out the report again by pressing
button.
Report is not saved, therefore when abandoning this level, it is deleted from the
scale volatile memory.
Press
button to abandon the window, the program returns to the main menu.
Press
button repeatedly to return to weighing.