ABL800 FLEX Reference Manual
4. User-defined corrections
Correction factors for oximetry parameters and bilirubin,
Continued
FCOHb
The following recommendations apply to
F
COHb:
Item Description
Units Fraction
Sample
The zero point (
F
COHb
≈
0) is saturated to approximately
SAT100, and
c
tHb is set to
≈
15 g/dL (9.3 mmol/L) and pH
≈
7.4.
Offset
±
0.050
The following recommendations apply to
F
MetHb:
FMetHb
Item Description
Units Fraction
Sample
The zero point (
F
MetHb
≈
0) is saturated to approximately
SAT100, and
c
tHb is set to
≈
15 g/dL (9.3 mmol/L) and pH
≈
7.4.
Offset
±
0.050
The following recommendations apply to
F
HbF:
FHbF
Item Description
Units Fraction
Sample
Radiometer recommends that
c
tHb in the adult samples (with
F
HbF = 0) and fetal samples (with high
F
HbF) is set to
≈
15
g/dL (9.3 mmol/L),
s
O
2
≈
100 %, and pH
≈
7.4.
The “Correction for HbF levels less than 20 %” function
should be enabled in order to have the
F
HbF value displayed
for the adult sample.
Averaging repeated measurements on blood from different
donors gives an optimized accuracy of the correction.
Averaging repeated measurements on blood from the same
donor also improves the accuracy.
Slope
0.800 - 1.200
Offset
±
0.20
Continued on next page
4-5