ABL800 FLEX Reference Manual
5. Performance characteristics
Performance test results - oximetry,
Continued
Test conditions
for oximetry
parameters
(continued)
Test
Description
Verification
6-10 ABL700 Series analyzers are tested over at least
2 days for all levels. Bias for the given mode is
calculated as difference compared to the C195
µ
L
mode.
Bias against the reference method is determined as
follows:
Bias = bias against C195 + C195 bias against
reference method.
The following parameters:
s
O
2
,
F
COHb,
F
MetHb and
F
O
2
Hb, are measured directly against the reference
built in the analyzer, and these parameters are
independent of the reference method.
Reduced verification
6 - 10 new analyzers are used over at least 1 day for
selected levels.
Bias for the tested mode is calculated as follows:
Bias = bias against C195 + C195 bias against
reference method.
Modes which are not tested are described as "N/A".
Simple verification
6-10 analyzers are tested at one extreme level over 1
day. Bias is not determined; bias values for the modes
with similar wet section programs are used.
The measuring modes were tested as follows:
Test Analyzer
Mode
Reference ABL735/25/15
C195
Verification
ABL735/25/15 S195, S95, S85, C95, C55, C35 MET,
C35 OXI
Continued on next page
5-31