5. Performance characteristics
Performance test results - oximetry,
Continued
FHHb -
micromodes
Bias:
FHHb
ABL835/25
ABL830/20
ctHb
(g/dL)
FHHb
(%)
S95 C95 S85 C55 C35 FM* C55 C35
15
0
0.09 N/A N/A 0.15 0.10 N/A N/A 0.10
* FM = FLEXMODE (no message)
Imprecision:
ctHb (g/dL)
FHHb (%)
S
0
S
D
S
ABL
S
X
15
0
0.05 0.10 0.30 0.35
FHbF – adult
blood
Bias (macromodes):
FHbF
ABL835
ABL830
FHbF (%)
ctHb (g/dL)
S195
FM*
S85
0 10 3.3 3.3
3.3
0 15 5.5 5.5
5.5
0 20 5.6 5.6
5.6
* FM = FLEXMODE (no message)
Imprecision (macromodes):
FHbF (%)
ctHb (g/dL)
sO
2
(%)
S
0
S
D
S
ABL
S
X
0 10 100
4
4
5
8
0 15 100
2
3
7
8
0 20
100
2
2
10
11
Continued on next page
5-37