ABL800 FLEX Reference Manual
5. Performance characteristics
Performance test results - oximetry,
Continued
FHbF – fetal
blood
(continued)
Bias (micromodes):
FHbF
ABL835
ABL830
ctHb
(g/dL)
FHbF
(%)
S95 C95 S85 C55 C35 FM* C55 C35
10
80 5.9 5.9 5.9 5.9 5.9 5.9 5.9 5.9
15
80 3.3 3.3 3.3 3.3 3.3 3.3 3.3 3.3
20
80 2.6 2.6 2.6 2.6 2.6 2.6 2.6 2.6
* FM = FLEXMODE (no message)
Imprecision (macromodes):
FHbF (%)
ctHb (g/dL)
sO
2
(%)
S
0
S
D
S
ABL
S
X
10
4 5 6 9
15
3 3 6 8
80
20
100
2 3 6 7
NOTES: a, b.
Contribution to
Imprecision
Specifications
from S7770
The following corrections should be geometrically added to S
Inst
and S
X
for the
analyzer's wavelength calibrated with the S7770:
Parameter Mode
Level
Correction
(percentage
point)
Macromode All
0
c
tHb
Micromode All
0
s
O
2
All
s
O
2
(100 %)
0.23
F
O
2
Hb All
F
O
2
Hb (100 %)
0.15
F
COHb
All
F
COHb (20 % and 0 %)
0.40
F
HHb All
F
HHb (0 %)
0.23
NOTES:
a.
pH = 7.4
±
0.1.
F
HbF is adjusted with the pH sensitivity to a nominal
pH=7.4. For further details please refer to the
Interference Tests
section for
oximetry parameters.
b. Specifications for imprecision are derived from worst-case values found
during internal laboratory tests. 40 % relative is then added as a safety
factor.
5-39