1. Potentiometric measuring principles
ABL800 FLEX Reference Manual
pCO
2
electrode,
Continued
Corrections -
blood samples
(continued)
B
= barometric pressure in kPa
p
H
2
O
= partial pressure of saturated water vapor (6.2751
kPa)
B
0
= correction cut-off
B
1
= correction factor
A
BL8XX
FLEX
Mode A
o
A
1
A
2
A
3
B
0
B
1
Eq.
35/25/15 S195
-0.003573
1.1126 0.0051
-0.0000002
A
S95
-0.003573
1.1126
0.0051
-0.0000002
1.000
-0.016
A,
B
S85
-0.003573
1.1126
0.0051
-0.0000002
A
C95
-0.003573
1.1126
0.0051
-0.0000002
1.013
0.010
A,
B
C55
-0.003573
1.1126
0.0051
-0.0000002
1.12
-0.28
A,
B
*FM
(no
message)
-0.003573 1.1126 0.0051 -0.0000002
A,
B
*FM
(message
874)
-0.003573 1.1126 0.0051 -0.0000002 1.013 0.010
A,
B
*FM
(message
873)
-0.003573 1.1126 0.0051 -0.0000002 1.013 0.010
A,
B
*FM
(message
872)
-0.003573 1.1126 0.0051 -0.0000002 1.0884 -0.1619 A,
B
*FM
(message
871)
-0.003573 1.1126 0.0051 -0.0000002 1.090 -0.150 A,
B
30/20/10 S85
-0.003573
1.1126 0.0051
-0.0000002
A
C 55
-0.003573 1.1126
0.0051
-0.0000002
1.12
-0.28
A, B
*FM
(no
message)
-0.003573 1.1126 0.0051 -0.0000002 1.013 0.050
A,
B
*FM
(message
872)
-0.003573 1.1126 0.0051 -0.0000002 1.0819 -0.0495 A,
B
*FM
(message
871)
-0.003573 1.1126 0.0051 -0.0000002 1.090 -0.150 A,
B
*FM = FLEXMODE
Continued on next page
1-18