ABL800 FLEX Reference Manual
5. Performance characteristics
Performance test results - oximetry,
Continued
FCOHb –
micromodes
(continued)
Bias:
FCOHb
ABL830/20
ctHb (g/dL)
sO
2
(%) FCOHb (%)
FM*
C55
C35
15 100 0
−
0.02
0.08 0.08
7 100
20
N/A
N/A
N/A
15 100 20
N/A
N/A
N/A
25 100 20
N/A
N/A
N/A
* FM = FLEXMODE (no message)
Imprecision:
ctHb (g/dL)
sO
2
(%)
FCOHb (%)
S
0
S
D
S
ABL
S
X
15 100 0
0.05
0.10
0.35
0.40
7 100
20
0.10
0.10
0.75
0.80
15 100
20
0.05
0.10
0.70
0.75
25 100
20
0.05
0.10
0.70
0.75
FMetHb -
macromodes
Bias:
FMetHb
ABL835/25
ABL830/20
ctHb (g/dL)
sO
2
(%)
FMetHb (%)
S195
FM*
S85
15 100 0 0.01
−
0.03
0.06
15 100 20 N/A
0.10 N/A
* FM = FLEXMODE (no message)
Imprecision:
ctHb (g/dL) sO
2
(%)
FMetHb (%)
S
0
S
D
S
ABL
S
X
15 100 0 0.10
0.10
0.25
0.30
15 100 20 0.05
0.10
0.35
0.40
Continued on next page
5-35
Содержание ABL800 FLEX
Страница 198: ...Traceability certificates...
Страница 199: ...ABL800 FLEX Reference Manual 7 Solutions and gas mixtures 7 9...
Страница 200: ......
Страница 201: ...ABL800 FLEX Reference Manual 7 Solutions and gas mixtures 7 11...
Страница 202: ......