5. Performance characteristics
ABL800 FLEX Reference Manual
Performance test results – ctHb
Primary
reference
method
HiCN method recommended by NCCLS
[
6
]
.
Bias
REF
The FLEXMODE on the ABL830/35 analyzers was tested:
ctHb (mmol/L)
Bias
REF
N
15 (SAT0)
0.32
145
7 (SAT100)
0.04
145
15 (SAT100)
0.37
145
25 (SAT100)
0.97
145
N = number of measurements on several analysers used for the test.
Bias
ABL
This bias is presented by the following chart:
Bias (%)
ABL735 (mmol/L)
Continued on next page