5. Performance characteristics
ABL800 FLEX Reference Manual
Definition of terms and test conditions,
Continued
Test conditions
Test conditions to determine bias
ABL
, repeatability and total variation for pH,
p
CO
2
,
p
O
2
,
c
Ca
2+
,
c
Cl
–
,
c
K
+
,
c
Na
+
,
c
Glu,
c
Lac,
c
tHb were as follows:
Item
Description
Reference analyzers
5 ABL735 with AutoCheck module were used as a
reference. The C195 mode was used as a reference for
all measured parameters.
Primary reference
methods
As specified for each parameter further in this chapter.
Analyzers and test
modes
5 ABL835, 3 ABL830, and 3 ABL805 were tested
over 11 days in the following modes:
•
Syringe: S195, S165, S95, S85
•
Capillary: FLEXMODE, C95, C85, C55, C35
OXI, C35 MET.
Blood samples
Heparinized blood samples from healthy, voluntary
donors.
11 Blood pools were prepared to cover test ranges for
all measured parameters.
Blood measurements
The measurements were performed by different
operators.
Calibration solution and
gases
All calibration solutions and gases used for the tests
are traceable to Primary Reference Standards.
Traceability certificates for the ABL800 FLEX
calibration solutions and gases are found at the end of
chapter 7: Solutions.
Experimental
conditions
Ambient temperature: 22 – 25
°
C
Relative humidity: 30 – 50 %.
NOTES:
•
The solutions used in the performance tests are those recommended by
Radiometer. Performances using other solutions cannot be guaranteed.
•
The performance tests are performed under conditions where the analyzers are
not influenced by electromagnetic fields.
5-4