3. Optical measuring principles
ABL800 FLEX Reference Manual
Measurement and corrections,
Continued
Corrections for
ctHb (continued)
ABL8XX
FLEX
Mode F
dil
Equation
30/20/10 S85
1.0050
A
C55
0.9220
A
C35
OXI
0.9570
A
*FM (no message)
0.9570
A
*FM (message 872)
0.9490
A
*FM (message 871)
0.9440
A
*FM (message 870)
0.9230
A
*FM (message 869)
0.9230
A
*FM = FLEXMODE
The uncorrected total bilirubin concentration,
c
tBil(sample), measured on capillary
or syringe samples is corrected as follows:
Corrections for
ctBil
Equation A:
dil
cuv
F
F
e)
tBil(sampl
corr)
e,
tBil(sampl
c
c
=
where:
c
tBil(sample,corr) = corrected
c
tBil
F
cuv
= Analyzer-dependent constant determined at tHb
calibrations and automatically stored by the analyzer
F
dil
= Analyzer dependent constant determined during tests
against the reference method, which corrects for
c
tBil
dilution in the different aspiration modes.