6. Parameters
ABL800 FLEX Reference Manual
Measured parameters,
Continued
pCO
2
(continued)
Measuring range
mmHg; torr: 5.0-250
kPa: 0.67-33.3
Reference range
mmHg: 35– 48 (m); 32-45 (f)
kPa: 4.67-6.40 (m); 4.27-6.00 (f)
Conversion of units
p
(kPa) =
(mmHg)
133322
.
0
p
×
=
0133322
.
×
p
(torr)
p
(mmHg) =
p
(torr) = 7 500638
.
×
p
(kPa)
pO
2
Is used for both blood and expired air samples.
Definition Partial
pressure
(or
tension) of oxygen in blood.
High and low
p
O
2
values of arterial blood indicate
blood hyperoxia and hypoxia, respectively.
Unit mmHg;
kPa;
torr
Measuring range
mmHg; torr: 0.0-800
kPa: 0.00-107
Reference range
mmHg: 83-108 (m, f)
kPa: 11.07-14.40 (m, f)
Conversion of units
p
(kPa) =
(mmHg)
133322
.
0
p
×
= 0133322
.
×
p
(torr)
p
(mmHg) =
p
(torr) = 7 500638
.
×
p
(kPa)
Baro
Definition
Ambient barometric pressure (
p
(amb)).
Unit mmHg;
kPa;
torr
Measuring range
mmHg; torr: 450-800
kPa: 60.0-106.7
Reference range
-
Conversion of units
p
(kPa) =
(mmHg)
133322
.
0
p
×
= 0133322
.
×
p
(torr)
p
(mmHg) =
p
(torr) = 7 500638
.
×
p
(kPa)
Continued on next page
6-6