ABL800 FLEX Reference Manual
7. Solutions and gas mixtures
Calibration solutions
S1820 and
S1830
Use:
For calibration of the pH, electrolyte and metabolite
electrodes.
Quantity: 200
mL
Composition:
Contains the following substances with the stated nominal
concentrations:
Solution
Substance
Concentration
(mmol/L)
S1820
K
+
Na
+
Ca
2+
Cl
−
c
Glu
c
Lac
buffer
4
145
1.25
102
10
4
Maintains a pH of 7.40
S1830
K
+
Na
+
Ca
2+
Cl
−
buffer
40
20
5
50
Maintains a pH of 6.9
The exact values are included in the bar code.
Additives:
Preservatives and surfactants.
Storage: At
2-25
o
C (36-77
o
F).
Stability:
Expiration date and Lot No. are printed on a label.
Stability in use:
4 weeks for S1820
8 weeks for S1830.
S7770 tHb
Use:
For calibration of the cuvette optical path length in the
ABL700 Series analyzers. The calibrated value can be
c
tHb,
c
tHb and
c
tBil, or
c
tBil depending on the analyzer version.
Quantity: 2
mL
Composition:
Salts, a buffer, preservative and a coloring agent.
Storage:
Keep in a dark place at 2 - 25
o
C (36 - 77
o
F).
After opening the solution must be used at once.
-
7-3
Содержание ABL800 FLEX
Страница 198: ...Traceability certificates...
Страница 199: ...ABL800 FLEX Reference Manual 7 Solutions and gas mixtures 7 9...
Страница 200: ......
Страница 201: ...ABL800 FLEX Reference Manual 7 Solutions and gas mixtures 7 11...
Страница 202: ......