20
RIDA qLine® autoBlot | User manual
3. The X axis will move to its home position; this consists of moving to the left hand side.
4. The carousel drive will rotate to its home position; this consists of ensuring position
1 is infront of the barcode reader.
5. The tray table will move to its home position; this consists of tilting the table.
6. The X and Y arm will move out to the tip waste trough, where the Z will perform
an eject action, followed by a return home by all axes.
If the above is successful, the master
controller checks will be ticked off and
followed by quick camera and barcode
reader checks. If all are successful, the
screen should resemble what is shown
in Figure 5. At this point, the application
can be entered by pressing start.
If anything goes wrong during initialisation,
a cross will be put next to the component
being initialised and a message will be
shown telling the user to contact technical
support. If anything goes wrong here
during the first time set-up, please contact
your service provider.
An example of a failure screen is shown
in Figure 6. Note you will not be allowed
to enter the main application if any errors
occur.
Figure 5:
The interface that is shown after
a fully successful instrument initialisation
routine
Figure 6:
The interface that is shown when
there are failures in the initialization routine