44
RIDA qLine® autoBlot | User manual
2. Load a cleaning solution (e.g. 4 %
bleach or 4 % SDS) into the wash
bottle and fastening the caps back
on correctly.
3. Place the wash bottle back inside
the instrument and close the door
4. Press Ok on the pop-up
Once the pop-up has been closed, the
instrument will start to prime the lines of
the wash pump with the cleaning solution,
followed by a short soaking period; the
task is reflected on screen, as shown in
Figure 42.
Once the instrument has finished soaking
the lines, a new pop-up will appear that
asks the user to load DI water into the wash
bottle as shown in Figure 43. At this point,
the user is expected to perform the
following:
1. Remove the wash bottle from
the instrument.
2. Discard any remaining cleaning
solution and rinse the wash bottle
with de-ionised water.
3. Fill the wash bottle with de-ionised
water and securely fasten the caps.
Figure 41:
The pop-up asking the user
to load a cleaning solution
Figure 42:
The screen that is shown while
the cleaning solution is being used