![Pyramid Technical Consultants F3200E Скачать руководство пользователя страница 96](http://html1.mh-extra.com/html/pyramid-technical-consultants/f3200e/f3200e_user-manual_1632602096.webp)
PSI System Controls and Diagnostics
F3200E User Manual
F3200E_UM_180726 Page 96 of 107
24.2 Using the A60 Recovery Utility
Reconnect the power and let the F3200E boot up. Start the PTC DiagnosticG2 software and
discover devices. The A60 Recovery will appear. Connect to this and the recover screen will
open. You can confirm you are connected to the correct device in case of doubt by looking at the
serial number.
Figure 77. A60 recovery screen.
Using the information on the screen, Pyramid will be able to tell you if any of the firmware is
corrupted. It is possible to load individual firmware files without going through the full update
process using the Erase and Update firmware button, and this may restore the function of your
F3200E.
After using the A60 Recovery utility, power down the F3200E, restore the original jumper setting
and refit the top cover