PSI System Controls and Diagnostics
F3200E User Manual
F3200E_UM_180726 Page 65 of 107
14
Serial Communications
Data is generated at too greater rate by the F3200E for a 115 kbps serial interface to be useful for
general use. However, it can be useful to check or change network settings or query the setup of
the unit for diagnostic purposes. The commands are ASCII and use the SCPI format. A
command that ends with a query solicits a response. A command without a query is an
instruction to the device.
The F3200E supports 115.2 kbps RS-232 ASCII communication. In the following example the
F3200E is asked to return its identity. The Ethernet IP address and mode is requested, then the IP
address is set to a new value.
Figure 54. RS-232 terminal session.
14.1 RS-485 connection example
RS-485 is used to extend a serial connection over a long distance. RS-485 is not currently
enabled in the F3200E, but may be added for specific customer needs. If you have a RS-232
serial port on the host PC, then an RS-232 to RS-485 converter is used at the host PC end. The
built-in RS-485 of the F3200E allows a direct connection at its end.
A commonly-used converter for the beam position monitor application is the MOXA TC100.
The Moxa TCC-80 is a low-cost alternative for less critical applications. The converter should
be configured for four wire (full duplex) RS-485 operation. The F3200E provides parallel
termination for transmit and receive. It is therefore optional whether you also terminate at the
TC100 end, but there is no harm if you do. The recommended Dip switch setting for the TC100
is:
Sw1
Sw2
Sw3
OFF
ON
ON
The recommended Dip switch setting for the TCC-80 is: