PSI System Controls and Diagnostics
F3200E User Manual
F3200E_UM_180726 Page 92 of 107
23
Controls and Indicators
23.1 Front panel controls
None.
23.2 Rear panel controls
23.2.1 Reset button
Momentary push-button that forces a warm reset of the on-board processor.
Holding the button in during the first few seconds of powering up forces the device to the default
IP address 192.168.100.20.
23.2.2 Address switch
16 position rotary switch setting device address for use when you have the F3200E as a slave on
a fiber optic loop. Choice of address is arbitrary, but each device in a fiber-optic loop system
must have a unique address. The switch is recessed behind the rear panel – use a fine flat-bladed
screwdriver to change the setting.
Setting
Function
0-15
Available address settings (x00 to x0F).
23.2.3 Mode switch
10 position rotary switch setting communications mode. This switch may be left at position 0 for
RS-232 ASCII communications.