PSI System Controls and Diagnostics
F3200E User Manual
F3200E_UM_180726 Page 51 of 107
11.8.3 Other
Figure 36. Other setup controls
Analog output 1 (pin 19 relative to pin 9 on the I/O connector) can be assigned to be manually
programmed, so that the output voltage is set directly from the host software on your PC, or to be
a monitor of the input currents. The currents that sum into the monitor signal are set on the
channels pane.
Various high voltage bias modules can be installed as a factory option. Internal jumper settings
identify the rating of the module to the internal software. A -2 kV module has been detected in
figure 36. The setpoint should be entered in volts including the polarity, for example “-1750”.
Pressing the HV button (
) enables the supply. The front panel LED will illuminate and you
should see a readback voltage similar to the setpoint. The High and Low settings allow soft
limits to be imposed on the high voltage settings.
The four general purpose analog inputs are typically used for relatively low bandwidth signals.
You can set the averaging period between 100 µsec and 1 second.
The actuator control turns on the 24 VDC output on the actuator connector (pin 1 relative to pin
2), intended for controlling pneumatic actuator solenoids. The Test (5V) output turns on 5 VDC
which is available on the actuator connector (pin 3 relative to pin 9) and on the signal input
connectors (pin 15 relative to pins 16-23).
11.9 Calibration panel
11.9.1 Current measurement calibration
The F3200E includes four high stability precision current sources which are used for the
automatic calibration process. Each source is assigned to one of the ranges.
Range
Calibration current
10 mA
8.333 mA
1 mA
833.3 µA