PSI System Controls and Diagnostics
F3200E User Manual
F3200E_UM_180726 Page 75 of 107
=
Separations indicate
beam velocity
Heights indicate
beam intensity
Shapes indicate
beam profile
Figure 61. Illustrative scanned beam example - analysis.
Figure 62. Burst of pulses captured in sweep mode, 20 sweep averaging.
16.4.2 Single shot acquisition at 1 MHz
Normal buffered acquisition is limited to a maximum sampling rate of 167 kHz to minimize
buffer overflow risk. However it is possible to use sweep mode to collect single shot data at 1
MHz for up to 1000 samples. Simply set the number of sweep averages to 1, and deliver a single
start trigger pulse. The F3200E will capture 1000 readings on each of 32 channels with 1 µsec
intervals, and you can save the data to a csv file if you are using the PTC DiagnosticG2 program.
Note that you must consider the analog bandwidth when looking at data that has been sampled at
this rate.