Using the Actuator Control Connector
Many detector devices are mounted on motion actuator systems so that they can be moved in and
out of the path of a beam of charged particles or ionizing radiation. The F3200E provides an
auxiliary input / output connector on the front panel intended for control of a pneumatic actuator,
but also available for other purposes.
Sw24 VDC power is available to drive a solenoid, relay coil or similar device. The
output is protected by a 130 mA automatically resetting fuse.
Two opto-isolated digital inputs are provided for limit switch sensing or general applications. 24
VDC power is available on pins 4 and 8, protected by 130 mA automatically resetting fuses.
Figure 74 shows a typical arrangement for pneumatic actuator control.
Solenoid
0V
+24 VDC
switched
+24 VDC
+24 VDC
Opto A input
Opto B input
1
5
6
9
Figure 74. Connecting a solenoid-operated actuator to the actuator control connector.
A relay sw5 VDC continuity test signal is routed to pin 3 on this connector in parallel to
pins on each of the signal input connectors. A FET-switched fast logic signal with 10 kohm
pullup to +5 V is provided on pin 6.
Figure 75 shows the F3200E internal circuits, to allow users to design other interfaces to the
connector as required.