Installation and setup
Operator´s Manual PiCCO
2
Page
Version 3.1
C-31
4.5
SpiderVision
The PULSION SpiderVision diagram shows all continuous parameters in dynamic conjunction.
Â
Selection of the SpiderVision diagram on the touch screen directly leads to individual configuration.
The brighter marked areas highlight the normal or target value
for the respective parameter.
The diagram is displayed
GREEN
as long as all displayed
parameters are within the normal or target value range.
The diagram is displayed
YELLOW
immediately one of the
displayed parameters goes outside the normal or target value
range.
If the diagram appears
RED
, two or more displayed
parameters are outside the normal or target value range.
Figure 33: "Spider" screen
All parameters which were given a target value are marked with the target value symbol
.
For details concerning the adjustment of normal values to target values please refer to chapter C5.7 Normal
and Target Values.