Installation and setup
Operator´s Manual PiCCO
2
Page
Version 3.1
C-11
CAUTION
When the PiCCO
2
is connected to a bedside monitor, perform a zero adjustment of the
PiCCO
2
before zeroing the bedside monitor.
3.3
Setup and start
3.3.1
Mains voltage
For connection to mains power only the supplied cable with protective connector may be used and connected
to a corresponding power socket with an intact protective earth conductor.
Ensure the connection to the potential equalization cable and switch on the device by pressing the mains
switch at the back of the device (position I). Connection to the mains power supply is indicated by the green
light at the ON/OFF key at the front of the device.
WARNING:
Do not reconnect the PiCCO
2
to electrical power if liquid has entered the device. A
short circuit may damage the device and cause hazardous conditions for patient and user.
The PiCCO
2
must always be connected to the protective earth conductor. Never use mains supply
cable or extension lines without protective earth conductor.
CAUTION:
If the connection to the protective earth conductor cannot be ensured, disconnect the
device from the mains supply and only use the device on battery power.
3.3.2
Switch on the device
Press the ON/OFF key for 1 second. The device needs a short activation time. This is indicated by the
blinking of all LEDs.
After having switched on the device, the PiCCO
2
automatically leads through the necessary screens,
depending on the connected patient cables.
During the start procedure, the
button will lead to the next screen.
While in actual operation each respective screen can be exited with the
button.
WARNING:
During the start up procedure an acoustic sound appears at the end of the activation
time. If this sound is not heard no acoustic alarm is available.
CAUTION:
If the device is not standing on a slip-proof surface and to prevent displacement, hold
the unit securely while the function keys are being pressed.
If the system check detects a failure, no function will be available and “SERVICE” is displayed on
the screen. Turn the PiCCO
2
off and contact your local PULSION representative. Do not attempt
to use or repair the PiCCO
2
.