Installation and setup
Operator´s Manual PiCCO
2
Page
Version 3.1
C-41
5.6.2
Configuration of central venous catheter
Setting of the individual injection site and injectate volume for the central venous catheter.
L
New in version 3.1:
Inj. Site:
V. Jug./Subcl. / V. Femoralis
Selection of the injection site. Default setting is V. Jug./Subcl.
Inj. Volume:
2ml / 3ml / 5ml / 10ml / 15ml / 20ml
Setting of the individually required injectate volume. The recommended injectate volume will be displayed
depending on body weight of the patient (see also Figure 40: "Measurement configuration" screen). Default
setting for an adult patient is 15ml.
5.6.3
Configuration of ScvO
2
measurement
Hb / Hct
Hb / Hct
Optional input of Hb or Hct. Default setting is Hb.
Hb Unit
g/dl / mmol/l
Input of Hb can be made in the preferred unit. Default setting is g/dl.
5.6.4
Configuration of ICG measurement
The ICG measurement option only appears if a LiMON module is connected.
ICG Dose
0.25mg/kg / 0.5mg/kg
Default setting is 0.5mg/kg
ICG quantity
mg
The necessary quantity of ICG, dependent on the settings and the patient´s body weight. The calculated value
may not be altered.