Installation and setup
Page C-10
Operator’s Manual PiCCO
2
Version 3.1
3.2.2
Central venous oxygen saturation
Disposables:
(3) Standard CVC
To be installed according to instructions for use. The distal lumen of the central venous catheter (CVC) has to
be designed for guide wires with Ø 0.032´´. Thermodilution injection, infusions and measurement of CVP
through the distal port is not recommended when a CeVOX probe is placed (see chapter C3.2.3)
(10) CeVOX fiberoptic probe:
Carefully place the probe through the distal lumen of the central venous catheter (CVC) according to the
instructions for use. Do not install a 3-way-stopcock between the probe and the hub of the distal lumen of the
central venous catheter (CVC). Remove the clamp on the CVC lumen in any case in order to avoid damage of
the probe.
Modules:
(11) CeVOX module (PC3010 / PC3015)
For connection of the CeVOX probe with the PiCCO
2
(SO
2
port)
(13) LiMON module (PC5100)
For connection of the LiMON reusable sensor with the PiCCO
2
(SO
2
port)
Sensors:
(12) LiMON reusable sensors (PC51100/PC51200/PC51300)
For connection of the patient to the LiMON module
3.2.3
Application of injectate temperature sensor housing and CeVOX probe at the same time
As the CeVOX probe is placed through the distal lumen of the central venous catheter, the next largest lumen
of the CVC catheter has to be used for injection of the thermodilution indicator and the central venous
pressure (CVP) measurement.
3.2.4
Transmission of continuous pressure to bedside monitor
The PiCCO
2
allows the transfer of continuous arterial and central venous pressure to all conventional bedside
monitors with standard excitation voltage of 5µV/V/mmHg.
(14) Adapter cable for pressure transfer to bedside monitor (PC85200)
For connection of the individual pressure transfer cable with the PiCCO
2
(
AP / CVP at rear side of device)
(15) Pressure cable to bedside monitor
Transmission of arterial pressure (AP) and central venous pressure (CVP) to patient bedside monitor. This
cable is available for all conventional patient monitors.
NOTE
Set the pressure input channel of bedside monitor to 5µV/V/mmHg. Other settings will lead to
incorrect blood pressure values being displayed on the bedside monitor.
Also set the pressure input channel of bedside monitor to 5µV/V/mmHg when using PULSION
monitoring kits directly connected to the bedside monitor.