Appendix
Operator´s Manual PiCCO
2
Page
Version 3.1
E-3
Measured Parameters:
Label
Unit
Lower limit
Upper limit
Accuracy
PCCO
l/min
0.25
25.0
Coeffici
ent of variation ≤ 2 %*
SV
ml
1
250
Coefficient of variation ≤ 2 %*
CO
l/min
0.25
25.0
Coefficient of variation ≤ 2 %*
GEDV
ml
40
4800
Coefficient of variation ≤ 3 %*
ITBV
ml
50
6000
Coefficient of variation ≤ 3 %*
EVLW
ml
10
5000
Coefficient of varia
tion ≤ 6 %*
ScvO
2
%
1
99
±3 % for 50
–
80 %
SpO
2
%
45
100
70% < SpO
2
< 100%:
≤ 2 %
SpO
2
< 70% not validated
PDR
%/min 0.0
50.0
±3 % / min (absolute) or 15 % relative
R15
%
0.0
99.9
n.a.
TB
°C
25
45
±0,1 °C
Tinj
°C
0
30
±0,1 °C
AP
mmHg 0
300
±4 mmHg (absolute) or 4 % relative
(acc. to EN 60601-2-34: 2000)
CVP
mmHg -10
50
±4 mmHg (absolute) or 4 % relative
(acc. to EN 60601-2-34: 2000)
*Coefficient of variation, measured using synthetic and/or database wave forms (laboratory testing)
Technical specifications are subject to change without further notice.