Installation and setup
Page C-46
Operator’s Manual PiCCO
2
Version 3.1
Implausible ScvO
2
value -
Recalibrate the monitor
Implausible value due to
defect or interference
Check probe, position and
optical module and repeat
calibration.
Error Optical Module -
Check/Replace optical module
Internal electronic problem in
optical module
If necessary replace
optical module
SpO
2
/ PDR
Error Message
Cause
Remedy
Error LiMON Sensor - Check /
Exchange LiMON sensor
Sensor error
(for more than 5 seconds)
Use a compatible and
correctly functioning
sensor
Error LiMON Module - Check /
Exchange LiMON module
Module error
(for more than 5 seconds)
Use a compatible and
correctly functioning
module
Error LiMON Sensor - Connect
LiMON sensor
Sensor not connected to
device or patient
(for more than 5 seconds)
Check connection of the
sensor
Pulse not detected - Relocate
LiMON sensor
Insufficient pulse signal
(for more than 5 seconds)
Reposition the sensor to a
place with better perfusion
Bad signal quality
–
Relocate
LiMON sensor
Insufficient pulse signal
(for more than 5 seconds)
Reposition the sensor to a
place with better perfusion
Too much ambient light
–
relocate
or cover LiMON sensor
Too much ambient light
(for more than 5 seconds)
Cover the sensor with
dark or non-transparent
material. Employ a
different sensor location
better protected from
ambient light
Strong movement artefacts
–
reduce LiMON sensor movement
Too much movement
(for more than 5 seconds)
Avoid patient movement
LiMON module disconnected
–
Check module connection
Module not connected to
device or patient
Check connection of the
module
Bad signal quality
–
Check signal
and repeat measurement
Low rate of perfusion
Relocate the sensor,
repeat the measurement