NOMEX
Service Manual - Cleaning
D909.131.00/02 en
89
Service Manual
WARNING
Improper execution of service tasks.
Hazards to Persons! Equipment Damage!
The following service tasks must only be per-
formed by specially trained staff.
Further service tasks may only be carried out
by PTW or by persons authorized by PTW.
16 Cleaning
WARNING
Liquids have entered the device.
Shock Hazard, Erroneous Measurement!
The device must be thoroughly inspected by
service personnel before being used again.
CAUTION
Disinfection of the surface with disinfectants on
a phenol base or peroxide compounds.
Equipment Damage!
Do
not
use disinfectants on a phenol base or
peroxide compounds to disinfect the surface.
NOMEX Multimeter Surfaces
•
Disconnect the NOMEX multimeter from the
PC before cleaning the surface.
•
Clean the NOMEX multimeter simply by wiping
it down with a moist cloth. Use water or a mild
soap solution. Liquids must not be allowed to
enter the device. Do
not
spray liquids onto the
device and connectors. The product cannot be
sterilized or disinfected.