NOMEX
Technical Manual - Technical Specifications
D909.131.00/02 en
83
Technical Manual
14 Technical Specifications
Only values with specified tolerances or limits are guaranteed. Values without tolerances are for informa-
tional purposes only.
14.1 NOMEX Multimeter
Device designation
NOMEX multimeter, type T11049 with
NOMEX software S030008
Manufacturer
PTW-Freiburg
Intended use
– diagnostic dosimetry system according to IEC 61674 for
absolute dosimetry and quality control measurements in X-
ray diagnostic radiology
– for acceptance testing and constancy checks in radiogra-
phy, fluoroscopy, dental X-ray, mammography, and com-
puted tomography (CT) applications
Operation mode
continuous operation
Measuring principle
measurement of the tube voltage of an X-ray unit, using semi-
conductor detectors and the dual-filter method
Measured quantities and units
dose [Gy] and dose rate [Gy/s]
exposure [R] and exposure rate [R/s]
dose per pulse [Gy/pulse] or [R/pulse]
pulses
pulse frequency [Hz]
exposure time [s]
mean peak voltage [kV], maximum peak voltage [kV] and
practical peak voltage [kV]
total filtration [mm Al]
half value layer [mm Al]
tube voltage waveform
dose rate waveform