NOMEX
Operating Manual - NOMEX Multimeter Setup
42
D909.131.00/02 en
6.2
Detector Menu
•
In the
Setup
menu, click the
Multime-
ter
line.
➔
The
Multimeter - Application
menu for the
NOMEX multimeter will be displayed:
Figure 15: Multimeter - Application menu
The following settings can be performed in the
Multimeter - Application
menu:
– selection of the NOMEX multimeter application
– display of the NOMEX multimeter calibration
and detector information.
6.2.1 Selecting the Application
The
Multimeter - Application
menu will be dis-
played.
•
Click the application that you want to select:
– RAD/FLU/DENT
– CT
– DENT-PAN
– MAM
➔
The selected application will be iden-
tified with a green check mark.
•
The MAM application requires further settings
to be performed. To access these settings,
click the
MAM
line.
•
Select or deselect the
Compression Paddle
function.
This setting ensures that the values
obtained by a measurement with a com-
pression paddle are corrected.
•
Click the beam quality you want to use:
– Mo / 30 µm Mo
– Mo / 0.5 mm Al
– Mo / 25 µm Rh
– W / 50 µm Rh
– Rh / 25 µm Rh
– W / 50 µm Ag
– W / 0.7 mm Al