Error Warnings (continued)
Error
Explanation
Warning: Startup Diagnostics:
Abort of previous run detected
Instrument detected that previous run was aborted. Check to see
whether plungers are loaded on the plunger bar. If so, run
Clean Up
from the 'Settings' Menu and follow prompts to safely remove them.
After plungers have been unloaded, remove the deck tray from the
instrument.
Warning: Startup Diagnostics:
Firmware version change detected
User warning informing user that a firmware version change has been
detected. Contact Promega Technical Services.
8.1 Use of a USB Flash Drive
• When using a USB flash drive, insert it before running any method and do not remove it until
the method is finished.
• Do not insert or remove a USB flash drive while instrument is running.
• Since USB flash drives may vary from vendor to vendor or from type to type, incompatibilities
can occur. If your USB flash drive is not detected after a few seconds or if you encounter any
issues with the USB flash drive, shut down and restart the tablet and the instrument, and try
a different brand of USB flash drive.
• If the tablet controller becomes unresponsive after inserting a USB flash drive, shut down and
restart the tablet and the instrument.
Troubleshooting
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...