46
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
Operating the Maxwell® CSC 48 Instrument
6.2 Starting a Method (continued)
Figure 34. Kit label indicating bar code to scan.
Shown in the red box is the bar code to
scan on the kit label for starting a purification run.
3. There are two potential error modes that can occur when scanning the bar code:
a. If the kit lot has passed the specified expiration date, a user prompt will appear explaining
that the kit has passed its expiration date and cannot be run. This kit should not be used,
and the user should select another kit that is within its specified expiration date to use for
the run.
b. If the scanned product catalog number has a purification method associated with it on
this instrument but the purification method is incompatible with the current software
mode, an information screen will appear indicating that the method is incompatible with
the current software mode. To run this method, shut down the Maxwell
®
CSC 48 RUO
software and double-tap the Maxwell
®
CSC 48 IVD software icon to use with the desired
purification kit.
c. If the product catalog number does not have a purification method associated with it on
this instrument, a user prompt will appear indicating that the software does not have a
method for this catalog number. Upon pressing the
OK
button on the user prompt, the
software will return to the ‘Home’ screen. Please contact Promega Technical Services
([email protected]) to get the most up-to-date information on the available
methods.
4. After the bar code, you will be presented with a ‘Cartridge Setup’ screen (Figure 35) that
allows you to specify the positions on the deck tray that will be occupied by cartridges and
to enter identification information for each sample. Each deck tray of 24 samples will be
displayed on a separate screen for sample entry. Toggle between the front (positions 1–24)
and back (positions 25–48) deck trays using the
Front
and
Back
buttons in the lower left
area of the screen.
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...