37
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
Preferences
Administrators can view preferences available in the Maxwell
®
CSC 48 IVD Mode software by
touching the
Preferences
button on the ‘Administrator Page’ screen. The ‘Preferences’ screen
displays a set of three tabs that can be used by the administrator to adapt the functionality of
the software to the needs of the laboratory (Figure 26).
Below are listed the tabs and associated preferences that can be set and a description of their
functions.
‘Common Settings’ tab
The ‘Common Settings’ tab on the ‘Preferences’ screen (Figure 26) provides administrators the
ability to specify the following options:
•
Allow deletion of results:
When checked, this box will allow administrators to delete
extraction report files from the local database of run reports.
•
Use Vision system:
The Maxwell
®
CSC 48 contains a Vision system camera that provides a
check of sample number, cartridge position and deck tray setup. Turning off the camera will
remove this functionality, relying solely on the user to manually specify number of cartridges,
cartridge positions and appropriateness of the deck tray setup (presence of cartridges at
specified deck tray positions, presence of plungers in cartridge well #8, presence of open
elution tubes). In the checked state, the Vision system camera function is on, while in the
unchecked state, the camera is turned off.
•
Auto Export:
Report files can be automatically exported to an administrator-defined location
at the end of instrument runs. Check the "Auto Export" box to enable this function, and then
touch the text box below this option to specify the path to which exported results files should
be saved. An ‘Export Folder’ screen will open with the current path specified in yellow
rectangles. Touch any aspect of the path to navigate to the desired folder. Touch the
Drive\
button to select the drive in which to search. Folders within the current path will be shown as
red rectangles in the main area of the screen. Results will be exported as tab-delimited text
files. If the options for Export to PDF and/or Export to Excel have been enabled, reports will
also be exported in the selected format(s) to the specified drive location.
When navigating away from the 'Preferences' screen, a prompt is displayed allowing the
administrator to save any changes that have been made. Touch the
Save
button to save any
changes and navigate away from the 'Preferences' screen. To leave the 'Preferences' screen
without saving changes, touch the
Don't Save
button. Touch the
Cancel
button to return to the
'Preferences' screen without saving changes.
Maxwell® CSC 48 User Interface
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...