45
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
Operating the Maxwell® CSC 48 Instrument
16964T
A
Figure 32. Maxwell
®
CSC 48 ‘Home’ screen.
Selecting the
Start
button begins the process
of running an extraction method on the Maxwell
®
CSC 48 Instrument.
2. You will be presented with the ‘Scan barcode’ screen (Figure 33) that will ask you to scan
the bar code on the kit box. The method bar code on the kit box is in the top right hand
corner of the Maxwell
®
CSC reagent kit label (Figure 34). Alternatively, touch the text box on
the ‘Scan barcode’ screen to manually enter the bar code information using the on-screen
keyboard. Maxwell
®
software requires manual entry of bar codes in the following format:
Product Catalog Number, Kit Lot Number, Expiration Date in year-month format (example:
AS17803221872018-05, where the product catalog number is AS1780, the kit lot number
is 322187 and the expiration date is 2018-05). Only if the bar code entered has this format
will the the
OK
button become active. The text to enter is shown to the right of the method
barcode in the top right corner of the Maxwell
®
CSC reagent kit label. Do not use the
information from the barcode at the bottom of the label. Once the bar code is correctly
entered, touch the
OK
button to get to the ‘Cartridge Setup’ screen. You can touch the
Back
button to return to the ‘Home’ screen. Scanning or entering the bar code
automatically selects the protocol to be run.
15623T
A
Figure 33. ‘Scan Barcode’ screen.
Scanning the method bar code automatically selects the
protocol on the Maxwell
®
CSC 48 Instrument.
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...