47
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
Note:
A red exclamation point icon is displayed at the top of cartridge positions when required
data is missing. Touch the red exclamation point icon for a description of the issue with a given
cartridge position.
a. Select cartridge positions that will be processed by touching the long rectangle for each
position that will be used. If the Vision system is activated on your system (see
Preferences in Section 5.3, Administrator Settings), the processing positions specified on
the deck trays as well as the correct setup of the deck trays are confirmed by the system
b. Once cartridge positions are selected, you must enter all sample tracking information
required by the administrator to proceed. Minimally required information is Sample ID and
Kit lot number. However, administrators may also require entry of cartridge bar codes,
elution tube bar codes, and up to two additional administrator-defined fields. All of the
required information must be entered for all selected cartridge positions before the
Proceed
button becomes active.
• Touch the black box below a cartridge position to scan or manually enter the
bar code information for the selected position (Figure 36). After a bar code is
scanned, the software will automatically move to the next empty bar code field
for a position. Once all sample identifier information for a cartridge is entered, the
software will automatically move to the next cartridge with empty bar code fields.
If cartridge and elution tube bar codes are required, we recommend that these are
scanned immediately before placing the cartridge or elution tube in the deck tray.
• Touch the Sample ID and additional required sample tracking information text
boxes to enter or scan the sample information.
• Multiple kit lots can be entered for a run by touching or swiping across multiple
black box positions and touching the Kit Lot Number text box to display a
‘Scan Barcode’ screen. Scan or enter the bar code information for the kit lot
that is being used for the selected cartridge positions. When entering multiple
kit lot numbers the bar code needs to conform to the format: Product Catalog
Number, Kit Lot Number, Expiration Date in year-month format (example:
AS17803221872018-05, where the product catalog number is AS1780, the
kit lot number is 322187 and the expiration date is 2018-05). The scanned bar
code will be checked to make sure it has the same product catalog number as
the originally scanned kit and that the kit is within the expiration date. An error
message is displayed if the kit is expired, if the kit does not match the selected
method, or if the kit is not supported by the instrument.
Note that positions with incomplete information will display a red circle with an
exclamation point at the top of the gray rectangle. The
Proceed
button will be gray
and inactive while required information is missing.
Operating the Maxwell® CSC 48 Instrument
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...