66
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
The Maxwell
®
CSC 48 Instrument has no user-serviceable parts and is designed to require
minimal maintenance. However, it is important to clean the instrument after every use. If samples
or reagents have been spilled, it is important to clean the instrument immediately to avoid
damage or contamination of samples.
Most parts of the Maxwell
®
CSC 48 Instrument have an anodized coating, which forms a
durable, easily cleaned barrier on the metal. Always turn off and unplug the instrument before
cleaning.
7.1 General Care
Wipe up any spills immediately. After each use, wipe off the magnetic rod assembly, plunger bar,
inside platform, and the outside of the instrument using cloths dampened with 70% ethanol. Do
not use other solvents or abrasive cleaners.
• Clean the Maxwell
®
CSC 48 Instrument after every use.
• Keep the vents in the back of the machine clear of dust.
• Do not remove the Maxwell
®
CSC 48 Instrument case for cleaning. This will void the
warranty.
• Do not use a spray bottle to soak instrument surfaces with large volumes of liquid.
• Never allow liquids to remain on instrument surfaces for extended periods of time.
• Keep all moisture away from the heated elution tube slots to prevent damaging the
heating elements.
Cleaning and Maintenance
7
Important.
Wear gloves
and appropriate
personal protective
equipment. If the
instrument is used
with biohazardous
materials, dispose
of any cleaning
materials used in
accordance with
your institutional
guidelines.
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...