26
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
5.2 User Interface Settings (continued)
Clean Up
Clean Up attempts to unload plungers from the instrument plunger bar if any have not been
appropriately unloaded after a method run. If your method run was aborted or had issues with
plunger loading or unloading, the Clean Up procedure should be run. Clean Up does not clean
the instrument and should not be confused with the Sanitization method.
You can run a Clean Up by touching the
Clean Up
button from the Maxwell
®
CSC 48 ‘Settings’
screen (Figure 17) [from the ‘Home’ screen touch the
Settings
button to access the ‘Settings’
screen]. Plungers can only be unloaded from the plunger bar if a cartridge is present beneath
the plunger to be unloaded. Place a cartridge (without a plunger) at each position on the deck
tray that did not unload a plunger properly during the run. On touching the
Clean Up
button,
you will be presented with a ‘Clean Up Checklist’ screen (Figure 18).
This screen presents you with a checklist of items that must be performed before attempting to
unload plungers.
• Cartridges are present at positions for which plungers are still present on the plunger bar
• Plungers are not present in any cartridges in the deck tray
• Deck tray has been placed in the instrument
You will have to select
Confirm
on every item of the checklist (Figure 22) before you can start
the clean up procedure. After you confirm that each of the checklist items has been performed,
touch the
Start
button to start the Clean Up process. The Maxwell
®
CSC 48 Instrument will
perform the Clean Up process and will generate and display the system report from the Clean
Up process. If plungers are not ejected after several clean up attempts, the operator should
contact Promega Technical Services ([email protected]; 1-800-356-9526) to determine
the next appropriate steps.
Maxwell® CSC 48 User Interface
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...