Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
6.2 Starting a Method (continued)
6. You will be presented with an ‘Extraction Checklist’ screen (Figure 38). This checklist
indicates the steps that need to be performed prior to starting an extraction process. You
must confirm that all checklist items have been correctly performed before the
Start
button
becomes active. Touch the
Cancel
button to return to the ‘Cartridge Setup’ screen.
• Sample preprocessing is complete. Sample preprocessing, if required, is described in the
Technical Manual for the specific Maxwell
®
CSC reagent kit.
• Place cartridges at the selected positions in the deck trays (Figure 39). Press down firmly to
snap the cartridges in place at both ends. There should be an audible click.
Figure 38. ‘Extraction Checklist’ screen.
This screen indicates the steps that must be
performed to prepare the deck trays prior to processing the selected samples. Individual
preprocessing steps are not indicated on the Extraction Checklist and should be performed prior
to this stage according to the technical manual for the kit being processed.
Figure 39. Placing the cartridges in the Deck Tray and pressing firmly to snap them in
place.
16979T
A
Operating the Maxwell® CSC 48 Instrument
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...