6.6 Sanitizing
Touch the
Sanitize
button on the 'Home' screen (Figure 56) to perform a UV sanitization of the
instrument. Make sure all samples have been removed from the instrument and all spills have
been cleaned prior to initiating the UV sanitization protocol. You will be presented with a
sanitization checklist (Figure 57) that informs you on how long the UV sanitization will take. It will
also ask you to confirm that no samples or eluates are present in the instrument prior to running
Sanitization. Once you confirm, the
Start
button is enabled. Touch the
Start
button to start UV
sanitization.
Figure 56. ‘Home’ screen.
Selecting the
Sanitize
button begins the process for UV treatment
of the Maxwell
®
CSC 48 Instrument.
Figure 57. ‘Sanitize Checklist’ screen.
The sanitization checklist indicates how long the UV
sanitization will take. Confirm that no samples or eluates are present in the instrument prior to
running Sanitization.
16964T
A
16989T
A
Operating the Maxwell® CSC 48 Instrument