33
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
16969T
A
Figure 23. Maxwell
®
CSC 48 ‘Sample Entry’ screen.
Use this screen to configure the bar
code information that must be entered for each sample prior to performing an extraction
method. Use the "Warn on duplicates" option to display a warning when multiple cartridges
within a run have identical Sample ID information. For any of the optional fields, it is possible to
specify whether the bar code information entered is required to match the Sample ID for each
processing position on the instrument.
3. The software can be configured to warn the user if duplicate bar codes are detected within
a run. The warning does not prevent the user from processing samples, but identifies
potential errors in bar code scanning for the user. To enable this feature, check the "Warn on
duplicates" checkbox.
4. After the Sample Entry settings are configured as desired, touch the
Save
button to save
these settings and return to the ‘Administrator Page’ screen.
Sanitization Settings
The Maxwell
®
CSC 48 Instrument contains a UV light that can help aid in sanitization of the
instrument. It is possible to specify the duration of the UV treatment and when UV light treatment
of the instrument should be performed.
1. From the ‘Administrator Page’ screen, touch the
Sanitization Settings
button to open the
‘Sanitization Settings’ screen.
Maxwell® CSC 48 User Interface
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...