4-24 PRODIGIT
OPP:START
(suitable for 3310C series)
Command syntax:
OPP:START {SP}{NR2}{;|NL}
Description:
The initial power of OPP testing.
OPP:STEP
(suitable for 3310C series)
Command syntax:
OPP:STEP {SP}{NR2}{;|NL}
Description:
The increment step power of each OPP test.
OPP:STOP
(suitable for 3310C series)
Command syntax:
OPP:STOP {SP}{NR2}{;|NL}
Description:
The maximun power of OPP test.
OPP:VTH
(suitable for 3310C series)
Command syntax:
OPP:VTH {SP}{NR2}{;|NL}
Description:
The threshold voltage when UUT output voltage is equal or lower than V-th during
OPP test.
TCONFIG NORMAL/OCP/OVP/OPP
(suitable for 3310C/3330A series)
Command syntax:
TCONFIG {SP}{NORMAL/OCP/OVPO/OPP}{;|NL}
Description:
There are 3 usefue test mode con be set normal mode .OCP mode (over current
protection) OVP mode.(over voltage protection ) OPP mode (over power
protection).
OPP mode is available in 3310c series only.
3. OVP mode is available in 3330A series only
Note
:
1.OCP/OPP and TCONFIG Function is available in 3310c REV
:
3.06 or later
Function is available in 3311c REV
:
3.06 or later
Function is available in 3312c REV
:
3.09 or later
Function is available in 3314c REV
:
3.07 or later
Function is available in 3315c REV
:
3.06 or later
Function is available in 3302c REV
:
2.9 or later