4-2 PRODIGIT
4.3 GPIB/RS-232C COMMAND LIST
GPIB/RS-232C setting and read back command are listed in table 4-1 and 4-2
。
Setting Preset numeric command
Model
Remark
3310A 3310C 3330A 3320
3250A
[
PRESet
:
]
BANK
{
SP
}{
d
}{
;
|NL
}
9
d=0~10
[
PRESet
:
]
WAVE
{
SP
}{
m
}{
;
|NL
}
9
M=0~4
[
PRESet
:
]
FREQuency
{
SP
}{
NR2
}{
;
|NL
}
9
40.0~70.0Hz
[
PRESet
:
]
RISE
{
SP
}{
NR2
}{
;
|NL
}
9
9
[
PRESet
:
]
FALL
{
SP
}{
NR2
}{
;
|NL
}
9
9
[
PRESet
:
]
SLEWrate
{
SP
}{
NR2
}{
;
|NL
}
9
[
PRESet
:
]
SLEWRATE{SP}
{
FAST|MIDD|SLOW
}{
;
|NL
}
9
[
PRESet
:
]
PREIod
:
{HIGH|LOW}
{
SP
}{
NR2
}{
;
|NL
}
9
9
9
[
PRESet
:
]
LDONv
{
SP
}{
NR2
}{
;
|NL
}
9
9
[
PRESet
:
]
LDOFfv
{
SP
}{
NR2
}{
;
|NL
}
9
9
[
PRESet
:
]
CC
{
SP
}{
NR2
}{
;
|NL
}
9
9
[
PRESet
:
]
CC
:
{
A|B
}{
SP
}{
NR2
}{
;
|NL
}
9
[
PRESet
:
]
CC
:
{
HIGH|LOW
}{
SP
}{
NR2
}{
;
|NL
}
9
9
9
[
PRESet
:
]
CP
:
{
HIGH|LOW
}{
SP
}{
NR2
}{
;
|NL
}
9
[
PRESet
:
]
CR
{
SP
}{
NR2
}{
;
|NL
}
9
[
PRESet
:
]
CR
:
{
A|B
}{
SP
}{
NR2
}{
;
|NL
}
9
[
PRESet
:
]
LIN
:
{
A|B
}{
SP
}{
NR2
}{
;
|NL
}
9
[
PRESet
:
]
CR
:
{
HIGH|LOW
}{
SP
}{
NR2
}{
;
|NL
}
9
9
[
PRESet
:
]
CV
:
{
HIGH|LOW
}
{
SP
}
{
NR2
}{
;
|NL
}
9
9
[
PRESet
:
]
CV
{
SP
}{
NR2
}{
;
|NL
}
9
[
PRESet
:
]
TCONFIG{SP} {NORMAL|OCP |OPP}{;|NL}
9
[
PRESet
:
]
TCONFIG{SP} {NORMAL|OCP|OVP }{;|NL}
9
[
PRESet
:
]
OCP:START {SP} {NR2}{;|NL}
9
9
[
PRESet
:
]
OCP:STEP {SP} {NR2}{;|NL}
9
9
[
PRESet
:
]
OCP:STOP {SP} {NR2}{;|NL}
9
9
[
PRESet
:
]
OCP:VTH {SP} {NR2}{;|NL}
9
9
[
PRESet
:
]
OPP:START {SP} {NR2}{;|NL}
9
[
PRESet
:
]
OPP:STEP {SP} {NR2}{;|NL}
9
[
PRESet
:
]
OPP:STOP {SP} {NR2}{;|NL}
9
[
PRESet
:
]
OPP:VTH {SP} {NR2}{;|NL}
9
Table 4-1 GPIB/RS-232 SETTING COMMAND SUMMARY
Note
:
1.OCP/OPP and TCONFIG Function is available in 3310c REV
:
3.06 or later
Function is available in 3311c REV
:
3.06 or later
Function is available in 3312c REV
:
3.09 or later
Function is available in 3314c REV
:
3.07 or later
Function is available in 3315c REV
:
3.06 or later
Function is available in 3302c REV
:
2.9 or later