3302C Series Mainframe Operation Manual 4-19
Description
:
The FALL load current slew rate of load level change or dynamic load can be
programmed by FALL command. The FALL slew rate of 3310 series Electronic
Load module can be fully independent to the RISE slew rate.
The FALL load current slew rate data must include decimal point, otherwise this
command is unable to execute. The effective FALL load current slew rate can be
programmed is the sixth digit after the decimal point. If the programming load
current level over the maximum specification at programmed range of 3310 series
load module, the fastest FALL slew rate will be sent to the load module.
Please make sure range I/II command before execute the load FALL slew rate
setting command, otherwise the 3310 series load module will adjust to fit the FALL
slew rate after programming the FALL command.
The engineering unit for FALL slew rate is
〝
A/us
〞
.
For Example
:
FALL 0.124 set FALL slew rate to 0.124 A/us.
SLEW Rate
Purpose:
Set the load current slew rate of 3330A series electronic load module.
Command Syntax:
SLEW{SP}{NR2}{;|NL}
Description:
Rise and Fall slew rate is the same from SLEW command setting of 3330A
electronic load module.
The Slew rate has two range, and it follows the CC mode range change
automatically, when CC Dynamic mode is in Range I, the slew rate is in range I, if
CC Dynamic mode is in range II, then Slew rate is in range II.
SLEWRATE
Purpose:
Set the load current slew rate of 3320 series electronic load module.
Command Syntax:
SLEWRATE{SP}{FAST|MIDD|SLOW}{;|NL}
Description:
This command selects the load current slew rate SLOW, MIDDLE and FAST.
For detail load current slew for SLOW, MIDDLE, FAST, and each model of 3320
series Electronic load module, please refer to the 3320 series Electronic load
manual.
For example
:
SLEWRATE FAST; selects the load current slew rate is fast.