3302C Series Mainframe Operation Manual 4-17
POWER Limit
Purpose
:
To set the upper/lower limit value of threshold power (W).
Command syntax
:
LIM:POW:{HIGH|LOW}{SP}{NR2}{;|NL}
Description
:
This command is to set the upper/lower limit value of threshold power (WATT).
When power (WATT) is lower than this lower limit value or higher than the upper
limit value, NG indicating light will come on to indicate
〝
NO GOOD
〞
.
For Example
:
LIN:POW:LOW 0.05, to set the lower limit value of threshold power (W) is 0.05 W.
LIM:POW:HIGH 250.0, to set the upper limit value of threshold power(W)is 250.0
W.
VA Limit
Purpose
:
To set the upper/lower limit value of threshold power (VA).
Command syntax
:
LIM:VA:{HIGH|LOW}{SP}{NR2}{;|NL}
Description
:
This command is to set the upper/lower limit value of threshold power (VA). When
power (VA) is lower than this lower limit value or higher than the upper limit Value,
NG indicating light will come on to indicate
〝
NO GOOD
〞
.
For Example
:
LIM:VA:LOW 0.05, to set the lower limit value of threshold power (VA) is 0.05 VA.
LIM:VA:HIGH 250.0, to set the upper limit value of threshold power(VA)is 250.0 VA.
PERIOD
Purpose
:
Set the Tlow/Thigh duration of dynamic load in Constant Current mode.
Command Syntax
:
PERI
:
{LOW|HIGH}{SP}{NR2}{NL}
Description
:
The PERiod of dynamic waveform is composed by Tlow and Thigh. The PERIOD
LOW and HIGH data must include decimal point, otherwise this command is
unable to execute.
The effective PERIOD LOW and HIGH can be programmed is the sixth digit a after
the decimal point. If the programming period of Tlow and Thigh setting over the
maximum specification at programmed range of load module, the maximum
duration of Tlow and Thigh will be sent to the load module.