4-20 PRODIGIT
SHORT ON/OFF
Purpose
:
Short the DC input of Electronic load.
Command syntax
:
SHOR{SP}{0|1|OFF|ON}{NL}
Description
:
This command applies the short across the input of the Electronic load. Executing
SHOR does not effect any programmed settings and the Electronic load will return
to them when the short is removed.
For Example
:
SHOR ON set the load module load input to short on state
SHOR OFF set the load module load input to short open state.
SENSE ON/OFF
Purpose
:
Set the voltage sense ON/OFF of Electronic load.
Command syntax
:
SENS{SP}{0|1|OFF|ON}{NL}
Description
:
This command applies to V-sense the input of the Electronic load.
For Example
:
SENS ON ; set the load module V-sense input to sense ON state. The voltage
input is from V-sense input connector of front panel.
SENS OFF ; set the load module V-sense input to sense OFF state. The voltage
input is from Load input connector of front panel.
DYNAMIC ON/OFF
Purpose
:
Set DYNamic ON or OFF command.
Command syntax
:
DYN{SP}{0|1|OFF|ON}{;|NL}
Description
:
DYN OFF set the load module in the 3302C mainframe to static load mode.
DYN 1 set the load module to dynamic load mode.
NG Enable /Disable
Purpose
:
Set Meter GO/NG check ON or OFF.
Command syntax
:
NGAB{SP}{OFF|ON}{;|NL}