4-22 PRODIGIT
When RANG command is executed, the values of the current level are adjusted as
follow:
1.If the existing current setting is within the new range, then the current level does
not change.
2.If the existing current level is setting in not within the new range, then the
current level is set to the maximum of new range. The maximum load current
level is shows in Table 4-5.
3.Please refer to Appendix C for proper programming procedure of 3310 series
electronic load module.
NAME
Purpose:
Setting the specifications of the 3320 series electronic load module.
Command Syntax:
NAME{SP}{3320|3321|3322|3324|3325}{;|NL}
Note:
This command is necessary for programming the 3320 series electronic load
module through 3302C mainframe.Send the NAME command before programming
other GPIB command.
Description
:
The configuration of 3302C mainframe is 3320.
NAME 3320
DUAL
Purpose:
Setting the Dual voltage or Dual current meter display of the 3332A series
electronic load module.
Command Syntax:
DUAL{SP}{DVM|DAM|OFF}{;|NL}
Description
:
The Dual V/A command is not in the control of CH A/B operation, it is an
independent command operation, this command is used for dual voltage meter or
current meter to be display on the two 5 digit LED display.
DUAL DVM; To set the meter is in dual voltage meter mode, the engineering unit is
"V".
DUAL DAM; To set the meter is in dual current meter mode, the engineering unit is
"A".
DUAL OFF; Disable the dual meter function, the upper 5 digit LED display is
voltage meter, the lower 5 digit LED display is current meter.