![Primes BeamMonitor BM+ 100S Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/primes/beammonitor-bm-100s/beammonitor-bm-100s_original-instructions-manual_1601729051.webp)
BeamMonitor BM+
35,0(6
51
Revision 02 EN - 02/2022
9.6.2
Advanced settings
1. Click on the
Advanced
tab.
2. Edit the options according to the ex-
planations in Tab. 9.6 on page 51.
Option
Explanation
Save device settings
All options marked with an asterisk in the
Device control
menu can be
saved in the EEPROM of the device.
These and other options for saving / loading configurations are de-
scribed in chapter 9.3.5 „Saving options“ on page 30.
Rotational speed in rpm *
For the BeamMonitor BM+, the rotation speed of the measuring tip is
always 1 562 rpm.
Calibrated wavelength(s) in nm *
The “calibrated wavelength” is the wavelength at which the device has
been validated. This is stored in the device and is displayed in the LDS.
Used wavelength in nm *
To calculate the beam quality factor M², the used wavelength must
be entered. Depending on the display in the option
Calibrated
wavelength(s) in nm,
the used wavelength can be entered in a defined
range.
For example, with a „calibrated wavelength“ of 1 064 nm, the used
wavelength can be entered from 1 000 - 1 100 nm.
Use one of the following options to set the wavelength of the laser used:
X
Enter a value in the input field.
X
Use the slider below the input field.
Move axes
With this option you can move to a defined y-position.
Use one of the following options:
X
Enter a value in the input field.
X
Use the slider below the input field.
This option is described in chapter 9.3.4 „Move axes“ on page 29.
Tab. 9.6:
Options in the
Device control > Advanced
tab of the
Linescan
measuring mode
Содержание BeamMonitor BM+ 100S
Страница 2: ......
Страница 3: ...BeamMonitor BM 35 0 6 3 Revision 02 EN 02 2022 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Страница 58: ...BeamMonitor BM 35 0 6 58 Revision 02 EN 02 2022 13 Declaration of conformity...
Страница 59: ...BeamMonitor BM 35 0 6 59 Revision 02 EN 02 2022...
Страница 68: ...BeamMonitor BM 35 0 6 68 Revision 02 EN 02 2022...